|
|
|
|
Integrated
Medicine |
|
Treatments
|
|
FAQ
Medical Microscopy
or Live Blood Analysis
What Is Medical
Microscopy?
Edgar Cayce, the seer of Virginia Beach, predicted in the
1930's that in the future the state of health of the individual would be
determinable on the basis of the evaluation of one drop of blood. This
time has arrived.
Through the advent of technological advances in microscopy, new
understandings and discoveries have been proven out by such leading
researchers as Royal Rife, Gaston Naessans,
Dr. Gunther Enderlein, and
many others. A new understanding of biology as a holistic science has
emerged. This new understanding is not only a key to recognizing
presently occurring imbalances in the body, it is also a most valuable
tool to implement for a preventive approach. Because the precursors to
serious health imbalances may be observed in the state of ever-present
floras found in the blood, health imbalances may be averted by reading
these early warning signs and making the necessary changes that will
allow one to rebalance the physiology. These markers are also applicable
in the course of tracking the progression and reversal of degenerative
conditions that may already be in motion. The observations of these
floras are made using what is known as a Darkfield microscope.
When utilizing today's conventional approaches to analyze blood, the
standard methods are to use either stains (which make certain factors in
the blood visible which would not be otherwise) or the electron
microscope (which provides ultrahigh magnifications). The limitation of
both of the preceding approaches is that the blood is effectively killed
through the processes utilized in observation. In a Darkfield, though,
the blood is stained with light frequencies, which allows the blood to
remain alive and active, thereby giving many otherwise unobtainable
clues as to the relative health of the blood, and thereby, the whole
organism. This is a very efficient biological probe available to the
integrative practitioner using a high-resolution projection microscope.
Analysis of both live and coagulated blood yields qualitative and
quantitative data on over 100 different risk factors and over 62
morphologies having distinct hormonal signatures in the blood.
Medical Microscopy or Live Blood Analysis is a cost-effective way to
assess the amount and general location of oxidative processes within the
body, as well as hormones, enzymes and other by-products of biological
stress. The generation of toxic oxygen products affects the body’s basic
cellular structures and biochemical pathways. This leads to further and
accelerated degeneration and specific diseases. These disease processes
will usually be in a chronic, progressive pattern such as
cancer,
cardiovascular diseases, diabetes, the collagen diseases and the
progressive neurological diseases.
Medical Microscopy or Live Blood Analysis can determine cellular
nutritional status. Nutritional status is essential to aid in curbing
and reversing the free radical cascade of destructive, deteriorating
cell structures. The interaction of the toxins given off by these cell
structures with normal body cells produces major systemic effects of
metabolic dysfunctions and challenges.
This unique assessment of peripheral blood can reveal a good prospectus
of the immune system. By observing the morphology of the white blood
cells and their activity in contrast to the extent of active foreign
antigens and microbes within the serum and red blood cells, the strength
of the immune system can be judged. Weakness and dysfunction in any one
of these three major components can influence the strength and function
of the other two. This can be indicative of a progressive disease
process.
Medical Microscopy or Live Blood Analysis cannot be depended upon for
specific diagnoses or pinpointing of specific organs involved.
However, good judgment can be made as to the general location of
pathological activity. Thus, this can direct the attending physician
toward organ complexes.
This unique analysis technique has the capability of offering the doctor
a quick and simple way to determine the extent of a disease process. It
helps determine whether it is functional or organic, acute or chronic,
mild or severe and gives a general idea as to what physiological pathway
of the body requires the greatest support. This can include oxidative
control, nutritional and enzyme correction and supplementation or immune
support.
The utilization of the variable projection microscope in analyzing the
serial drops of blood, air dried and coagulated in conjunction with High
Resolution Blood Morphology, is a simple, proven way to triage the new
patient, saving money and time as well as pain and suffering. It is an
intelligent way to direct further laboratory studies, physiological
and/or psychological examinations in determining a more accurate
diagnosis. Treatment can be directed toward normalizing physiological
functions by supporting known immune and nutritional deficiencies.
Integrative modalities are employed in eliminating microbial and/or
tumorfractions, which may become organized, walled-off and protected
from the body’s natural immune modalities.
Medical Microscopy or Live Blood Analysis lends itself readily to
monitoring physiological and morphology changes during treatment. This
enables Dr. Eddy to fine-tune the supportive therapies and know when
further support is no longer needed. This monitoring tells the physician
when the body’s basic cellular structures and biochemical pathways have
regained their integrity and the organ reserve has again been
reestablished. This, indeed, is truly demonstrating “state-of-the-art”
integrative medicine.
50 Most Frequently
Asked Questions About Optical Microscopy Answered by
Mortimer Abramowitz
Technical Consultant at Olympus America, Inc.
Question 1. What does the inscription 0.17 on the
objective signify?
Answer: 0.17 refers to the thickness, in millimeters, of
the cover glass that was assumed by the lens designer in
computing the corrections for the objective. For
objectives with a numerical aperture higher than 0.45, a
departure in this assumed thickness (or no cover glass
at all) may result in deterioration of the image.
Question 2. What is the significance of the N.A.
number inscribed on the outer barrel of an objective?
Answer: This number is the numerical aperture of the
objective, a measure of the light-gathering capacity of
the objective. The higher the numerical aperture, other
things being equal, the better the objective is able to
separate the details of the specimen in forming an
image; also the brighter the image. Higher N.A.
objectives are usually more expensive. The higher the
numerical aperture of the objective, the shallower the
depth of field (see definition of depth of field).
Question 3. What does the objective inscription 160
mean?
Answer: This number identifies a finite tube length
objective. 160 millimeters is the distance from the
opening of the nosepiece (where the objective is screwed
in) to the top of the observation tube (where the
eyepiece is inserted). If this distance is lengthened,
e.g. by insertion of accessories in the light path above
the nosepiece, spherical aberration will result, unless
optically corrected lenses are included in the
accessory.
Question 4. What does the objective inscription
signify?
Answer: This inscription identifies the objective as an
infinity-corrected objective. Light rays emerging from
such an objective are in parallel bundles projected
toward infinity. Such an objective, with its many
advantages, requires a tube lens in the light path to
converge the parallel rays so that they come to focus at
the plane of the eyepiece diaphragm.
Question 5. Some objectives have the inscription
Plan. What does that mean?
Answer: Plan designates an objective that projects, at
the eyepiece diaphragm plane, an image which is flat
from edge to edge of the field of view. Some objectives,
e.g. the new planapochromats and new planfluorites, will
give flat images even with eyepieces of field numbers (F.N.
see below) up to 26.5; others up to F.N. 22.
Question 6. If an objective carries the inscription
Planapo, what does that mean?
Answer: The term Planapo signifies a planapochromat, an
objective of the highest correction, corrected for four
colors chromatically and spherically. Such an objective,
for its magnification, will have a higher numerical
aperture than objectives of lesser correction.
Planapochromats are the best objectives for critical
resolution and color photomicrography. Other things
being equal, they usually have shallower depth of field.
They are also more expensive.
Question 7. What does the inscription PlanFl mean?
Answer: PlanFl denotes an objective that is a
planfluorite also called a plan-semi-apochromat. (Some
manufacturers call such objectives Fluars or Neofluars)
These objectives are also corrected for four
wavelengths, but not quite as completely as
planapochromats. Planfluorites are also well-suited for
color photomicrography and are less expensive than
planapochromats.
Question 8. Some objectives have no inscription about
their corrections or only the inscription Plan. What
kinds are these?
Answer: Such objectives are achromats or planachromats.
They are now corrected for three wavelengths
chromatically, and one or two wavelengths spherically.
They give their best images in green light. However, in
white light, the planachromats will yield satisfactory
images for color photomicrography but not as good as
objectives of better correction.
Question 9. Objectives usually have a color ring
inscribed. What do these colors signify?
Answer: The 4x or 5x has a red ring; the 10x yellow; the
20x green; the 40x or 50x or 60x blue; the 100x white.
These rings make it easier to visually identify the
magnification of the objective; the colors are standard
for most manufacturers. In addition, Olympus phase
objectives are further identified by having a red ring
nearer to the front lens of the objective (LB series) or
all inscriptions in green for the infinity series of
phase objectives.
Question 10. What is a No-cover glass or NCG or NC
objective?
Answer: Such an objective has been designed to look at a
specimen which is not covered by a cover glass; e.g. a
smear. At numerical apertures above 0.45, such an
objective will yield images free of spherical aberration
when smears or other uncovered specimens are examined.
Metallurgical objectives are almost always designed for
looking at uncovered objects; e.g. polished metals,
wafers, etc.
Question 11. Why would one use an objective marked
LWD or ULWD?
Answer: The letters identify a long working or
ultra-long working distance objective. The vertical
distance from the front lens of the objective to the
focused specimen (working distance) is much longer than
that for a similar magnification standard objective.
Such objectives are invaluable for looking up through a
culture vessel or Petri dish in inverted biological
microscopy; or examining IC wafers to prevent
inadvertent contact with the wafer; or inspecting solder
connections of mounted chips.
Question 12. Some objectives are marked NIC or DIC.
Why?
Answer: This designates an objective that is preferred
for use in Nomarski (NIC) or differential interference
microscopy (DIC).
Question 13. What does the inscription oil or oel or
WI on the objective mean?
Answer: Such an objective is an immersion objective,
requiring a drop of immersion oil (or water) between the
front lens of the objective in contact with the cover
glass or top of a smear. Unless there is oil contact
with such an objective, the image will be very poor.
Some manufacturers inscribe a black ring on the lower
part of the barrel to enable the user to instantly
recognize the need for oil contact. Similarly, the
letters WI refer to an objective that requires water,
rather than oil, as the immersion contact medium. To
achieve a numerical aperture of 1.0 or above requires an
immersion objective. WI objectives are especially useful
in the observation of living biological specimens.
Question 14. Why are some objectives marked UV?
Answer: Ordinary glass is relatively opaque to
ultraviolet light wave lengths below 400 nanometers. UV
objectives contain specially formulated glass elements
and coatings to transmit a relatively high percentage of
light of such wave lengths. These objectives are very
useful for near ultraviolet excitation in reflected
light fluorescence work. The new Olympus
infinity-corrected U-planapochromats and U-planfluorites
have such improved transmission in the near ultraviolet.
Question 15. What do the letters PL or NH mean on a
phase objective?
Answer: The letters PL stand for positive low, a phase
contrast in which the specimen appears darker than the
background of the field of view. Less commonly used, the
letter NH stand for negative high, a type of phase
contrast in which the specimen appears lighter than the
background.
Question 16. Why are some objectives marked P or POL
or SF?
Answer: These letters designate the objective as being
relatively strain-free (that is, the objective itself
has little or no effect on polarized light). Such
objectives are required for high quality polarized light
qualitative and/or quantitative investigations.
Question 17. Eyepieces, in addition to having the
magnification inscribed, are marked 20, or 22, or 26.5.
What do these numbers mean?
Answer: The designation is the field number of the
eyepiece. The higher the field number of the eyepiece
being used with a particular objective, the more
specimen area will appear in the field of view. The
diameter of the field of view, in millimeters, is
calculated by dividing the field number of the eyepiece
by the magnification of the objective. (For example,
with a 10x objective and an eyepiece with a field number
of 22, the diameter of the field of view would be 2.2
millimeters) For many microscopists, e.g. hematologists,
it saves time to be able to see more of the specimen at
a given time. Eyepieces with a field number of 26.5 are
called super-wide eyepieces.
Question 18. What do the inscriptions C or K or WF or
H on the eyepiece mean?
Answer: C or k identify a compensating eyepiece. Some
microscope objectives do not include correction for
lateral chromatic aberration. For such objectives
(Olympus LB series), the compensating eyepiece completes
the correction. WF means widefield; more of the specimen
to be seen at a given time. H signifies high eyepoint
which means that the user's eyes do not have to be
placed very close to the top lens of the eyepiece during
observation; a particular boon to eyeglass wearers.
Question 19. Why do some observation tubes have
notches cut into them.
Answer: The notch is meant for placement of the
projecting "locator" pin on eyepieces which have a
reticle installed. Such an eyepiece can be focused by
rotating its diopter adjust upper lens, while the
"locator" pin keeps the reticle (e.g. crosshairs or
micrometer scale) properly oriented.
Question 20. What are the outer dimensions of the
most common observation tubes?
Answer: These are usually 25 millimeters or 30
millimeters (as on the latest Olympus B-max microscopes
and on all Olympus super-wide observation tubes).
Question 21. What is a photoeyepiece?
Answer: This is an eyepiece meant for photomicrography,
not to be used for observation. The eyepiece "picks up"
the image projected by the objective, and projects that
image onto film plane inside a camera. Photoeyepieces
(also called projection lenses) usually come with low
magnification power because the images they project onto
film are often subsequently further enlarged (to lessen
chances of "empty" magnification).
Question 22. Some objectives, 20x magnification or
higher, have a spring-loaded or retractable front lens
assembly. Why?
Answer: Such objectives have a very short free working
distance; hence the danger of inadvertently crashing the
front of the objective into the cover glass or specimen.
A spring-loaded front lens assembly allows this part to
retract upon gentle pressure contact with a specimen. It
will NOT protect against rough, continuous pressure
contact.
Question 23. Some objectives come with a built-in
iris diaphragm. What is the use of this diaphragm?
Answer: The diaphragm is partially closed down during
darkfield microscopy in order to reduce the numerical
aperture of the objective below the lower of the two
numerical apertures inscribed on a darkfield condenser.
This action preserves the darkness of the background in
darkfield observation. The iris diaphragm is absolutely
necessity for high numerical aperture oil immersion
objectives (above N.A. 1.2) when using an oil immersion
darkfield condenser. For ordinary brightfield
observation, this iris diaphragm is left wide open.
Question 24. What is the purpose of the correction
collar found on some "dry" 20x, 40x, or 60x objectives?
Answer: The correction collar, when rotated, separates
or brings together some of the internal glass elements
of the objective. This action can correct for incorrect
cover glass thickness. In upright microscopes, the range
of correction of the collar is usually from 0.11-0.22
millimeters. For inverted microscopes, the range of
correction capability is from 0 (uncovered) to 2
millimeters to correct for thick culture vessels. The
purpose is to eliminate spherical aberration.
Question 25. What is meant by the free working
distance of an objective?
Answer: This is the vertical distance in millimeters, or
a decimal fraction of a millimeter, from the front of
the objective to the cover glass or uncovered specimen,
when the specimen is in focus. High magnification
objectives customarily have very short working
distances. The manufacturer should be able to supply you
with working distance data for each objective.
Question 26. What is meant by the adjustment distance
or parfocalizing distance of a microscope?
Answer: In most modern microscopes, this distance (45
millimeters) is the chosen distance from the opening of
the nosepiece to the focused specimen. By having a
standard distance for all objectives, it is easy to
rotate each objective, in turn, into the light path with
a minimum of refocusing. Such a battery of objectives,
despite various lengths of the objectives themselves, is
described as being parfocal. If, in addition, a focused
feature which has been centered in the field of view
remains centered as objectives are changed, the
objectives are described as parcentric.
Question 27. What is meant by resolving power of an
objective and how is that distinguished from resolution?
Answer: Resolving power of an objective refers to the
ABILITY of that objective to yield an image which
clearly separates points or lines lying close together
in the specimen. The shorter the distance (between point
or lines) the better the resolving power of the
objective. Resolving power is related to the numerical
aperture of the objective; the higher the numerical
aperture, the better the resolving power. The formula is
expressed as:
d = / 2N.A.
or
d = 0.61 / N.A.
The former is according to Abbe, the latter according to
Rayleigh. In the formula, d is the distance between two
closely lying points; is the wavelength of light being
used; N.A. is numerical aperture. Resolution is the
ACTUAL separation achieved by the microscope system. For
example, in Köhler illumination, the condenser diaphragm
is usually closed down somewhat in order to effect a
compromise between resolving power and contrast--to
enhance visibility.
Question 28. What is the highest numerical aperture
for a "dry" objective (which requires air between its
front lens and the specimen?)
Answer: The highest N.A. for a "dry" objective is 0.95.
If such an objective is meant to look through a cover
glass, it must have a correction collar. If such an
objective is designed for uncovered objects, e.g. smears
or metallurgical, it will not require a correction
collar.
Question 29. What is the relationship between
numerical aperture and brightness of an image? Between
magnification and brightness of an image?
Answer: Other things being equal, the brightness of an
image varies directly as the square of the N.A.; as the
fourth power in reflected light fluorescence. Thus,
higher N.A. yields brighter images. Conversely,
brightness varies INVERSELY as the square of the
magnification. Thus, other things being equal, higher
magnification will reduce the brightness of the image.
Question 30. Can I use an infinity corrected
objective on a finite tube length microscope?
Answer: No, because the finite system does not include a
tube lens to bring the parallel rays to focus.
Question 31. Can I use a finite tube length objective
on an infinity-corrected system?
Answer: Although you may be able to screw the objective
into the nosepiece, the presence of a tube lens in the
light path will result in a deteriorated image.
Question 32. Can I use an infinity-corrected
objective from another microscope manufacturer on an
Olympus infinity-corrected system?
Answer: No, that would not be advisable. The focal
length of Olympus' tube lens (180mm) is not the same as
that of other manufacturers. As a result, the
magnification of the objective would not be accurate.
Also, aberrations would be introduced because other
companies correct for lateral chromatic aberration in
the tube lens; the new Olympus objectives for the B-Max
series achieve this correction in the objectives
themselves. Also, the objective probably would not be
parfocal with Olympus objectives.
Question 33. Can the new series Olympus
infinity-corrected metallurgical objectives be useful on
a biological B-Max microscope?
Answer: Yes. For covered or uncovered specimens,
metallurgical objectives of N.A. 0.40 or less should
give satisfactory images. For an N.A. above 0.40, the
specimen should not be covered by a cover glass, since
the metallurgical objectives are corrected for uncovered
specimens. The working distance may be very short, and
the parfocality with biological objectives may not be as
accurate.
Question 34. Can a planachromat objective be useable
in reflected light fluorescence?
Answer: The answer is a provisional yes. Planachromats
may serve satisfactorily for blue or green excitation
waves lengths. However, the planachromat's glass
elements may themselves fluoresce in excitation in the
near ultraviolet. Also, planachromats, for their
respective magnifications, have lower N.A.'s than do
planfluorites or planapochromats-thus the images may be
not as bright.
Question 35. Can phase contrast objectives be used
for regular brightfield observation?
Answer: Yes. The phase condenser should be moved to the
O position and standard Koehler illumination procedure
should be employed. The brightfield images will be
nearly as good as if a regular brightfield objective
were being used.
Question 36. What objectives are generally chosen for
clinical laboratory microscopes?
Answer: Such microscopes can function satisfactorily
with ordinary achromats or planachromats. For research,
or for best color photomicrography, planfluorites or
planapochromats are preferable (and more expensive).
Question 37. When a 40x objective is used, the image
may appear worse than with a 20x objective. Why?
Answer: The specimen may be covered with a cover glass
much thicker than the standard 0.17 millimeters or may
be mounted in a thick mounting medium under the cover
glass. The solution could be to use a "dry" objective
with a correction collar; or to substitute a 40x or 50x
oil immersion objective for the 40x "dry" lens since the
immersion objective will be less sensitive to variations
in cover glass thickness.
Question 38. With an objective of a given
magnification, why shouldn't one use increasingly higher
magnification eyepieces to achieve higher total
magnification?
Answer: To maintain useful magnification, that is,
magnification yielding satisfactory clarity and
resolution, one must avoid making the specimen appear
bigger but not clearer ("empty" magnification). The
general rule of thumb in microscopy is that the total
magnification should probably not exceed 750x to 1000x
the numerical aperture. For example, with a 40x, N.A.
0.65 objective, the total magnification (The multiple of
the eyepiece and the objective magnifications) should be
between 480x and 650x.
Question 39. What is meant by depth of field?
Answer: This is the vertical distance in the specimen,
usually in microns, measured from above and below the
exact plane of focus which still yields an acceptable
image. The higher N.A. the shallower this distance.
Question 40. What is the depth of focus?
Answer: This is the vertical distance in the IMAGE space
(at film plane), from above and below the exact image
plane, that appears satisfactorily sharp. Contrary to
what might be expected, the depth of focus is shallower
for low magnification objectives. This makes it more
difficult to get sharp images, in photomicrography, with
low power objectives.
Question 41. What is the value of a green filter
placed in the light path?
Answer: Because achromats and planachromats are best
corrected spherically for green light, and since the use
of monochromatic light eliminates chromatic aberration,
the performance of achromats is markedly improved with
the use of a green filter. Also, phase contrast
objectives are computed to give best phase images in
green light.
Question 42. Some microscopes come equipped with a
so-called daylight blue filter. What is this used (or
misused) for?
Answer: The "daylight blue" filter is a filter for
observation only. It furnishes a pleasant pale blue-gray
background to the field of view. It is NOT meant for
photomicrography with daylight color film. Such film
requires a blue conversion filter such as the Olympus
LBD or Kodak 80A filter. The conversion filter boosts
the color temperature of the light source, thus
simulating light of daylight color temperature quality
(5500 degrees Kelvin) required for daylight balanced
color film.
Question 43. How does a substage condenser function
in helping to provide excellent images?
Answer: The working numerical aperture of a microscope
is the sum of the N.A. of the objective plus the N.A. of
the condenser divided by 2. In order to retain more good
contrast with good resolution, the condenser iris
(aperture diaphragm) is usually opened to about 3/4 of
the N.A. of the objective. Condensers also vary in their
color and spherical correction-from the relatively
modest correction of an Abbe condenser to the highest
correction aplanat-achromat condenser. The
aplanat-achromat condenser, with its high N.A., is the
best choice for color photomicrography. Some condensers
have a swing-out upper element. With this element out of
the light path, the condenser is able to fill the field
of view, without vignetting, of a 4x or a 2x objective.
Question 44. Some nosepieces have wider diameter
openings into which the objectives are screwed. Why?
Answer: In metallurgical microscopy, objectives are
available for reflected light brightfield and/or
darkfield observation. These objectives have a wide
diameter barrel for use in darkfield reflected light;
hence need for wider openings of the nosepiece. Such
objectives may be labeled Neo or B/D or BF/DF.
Question 45. How can you minimize the likelihood of
immersion oil getting on the 40x "dry" objective?
Answer: One expedient often used in clinical
laboratories is to mount the "dry" 40x on the opposite
side of the nosepiece from the 100x oil immersion
objective. This mounting arrangement reduces the
likelihood of inadvertent dipping of the 40x "dry"
objective into immersion oil as you rotate the nosepiece
between the 100x oil objective and the 40x "dry"
objective.
Question 46. Is there a good alternative choice for
the 40x "dry" objective to avoid the oiling "hazard"
described?
Answer: If the 100x oil objective is used frequently, it
might be advisable to substitute a 50x OIL immersion
objective in place of the 40x "dry" objective. The 50x (N.A.
0.90) oil planachromat will yield much brighter images,
with better resolution than the standard 40x (N.A. 0.65)
dry planachromat or achromat.
Question 47. Are there disadvantages to choosing a
50x oil objective as suggested?
Answer: The 50x oil planachromat is approximately twice
as expensive as the 40x "dry" planachromat. Also, oil
objectives are difficult to use with haemocytometers
because the oil may adhere and inadvertently lift off
the coverslip.
Question 48. Is it usually advisable to buy the
"best" (highest correction) objectives that your budget
will allow?
Answer: Yes, but with a few caveats: If you are doing
mostly observations of specimens thicker than several
microns, planachromats or planfluorites may serve quite
well because they have greater depth of field than
comparable magnification planapochromats. For color
photomicrography, planfluorites are capable of rendering
better color images than can planachromats. For the
finest color photomicrography (and observation) of
minute details, planapochromats are the best choice but
several times more expensive than the planfluorites.
Question 49. Is it important to choose the highest
available N.A. objective for video microscopy of minute
specimen details?
Answer: Yes. Although the image observed through the
eyepiece, with the condenser aperture 100% open, may
have so much glare that details become invisible, the
"information" probably is there. Video enhancement
techniques, controlling brightness and contrast, can
process this "information" and render and excellent,
high resolution, visible video image of the details.
Question 50. Are the NEW Olympus infinity-corrected
metallurgical objectives parfocal with the NEW
infinity-corrected biological objectives?
Answer: Yes, they are.

Related Site:
Treatments
Treatments Programs:
|
|
 |
Integrated Medicine
combines Western
medicine with Complementary and Alternative medicine
and mind-body-spirit approaches to health and
healing.
Live Blood Analysis
Two drops
of blood under a specialized high powered ultra-dark
field microscope, reveals anomalies in the blood.
The unique
tool for prevention.
Ozone-Oxygen-Therapy
is recognized by most as
the most powerful and versatile therapy known in
alternative health because it plays a vital role in
maintaining the well-being of the body.
Check it
out why.
Contact the Doctor

contact the doctor
Disclaimer
This information is
provided for general medical education purposes only and
is not meant to substitute for the independent medical
judgment of a physician relative to diagnostic and
treatment options of a specific patient's medical
condition.
In no event will The DrEddyClinic.com be liable for any
decision made or action taken in reliance upon the
information provided through this web site.
|
|
|
DrEddyClinic.com
Chiang Mai 50230, Thailand
Phone.
+66-53-436284
Fax. +66-53-436284
Mobile. 098505066
email
contact
|
 |
 |
 |
|
|